Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Tegam 135A: CAL ADJ /5500 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 98 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # 135v.bmp # 135adj1.bmp # 135adj2.bmp # 135adj3.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: 0 - 50degC 1.004 DISP Humidity: < 80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP [32] WARNING 1.006 DISP 1.006 DISP To prevent a shock hazard, remove all test leads 1.006 DISP from the instrument and turn the instrument OFF 1.006 DISP before disassembling the UUT. 1.006 DISP 1.006 DISP [32] WARNING 1.007 HEAD DISASSEMBLING INSTRUCTIONS 1.008 DISP Place the unit face down and remove the battery. 1.008 DISP 1.008 DISP Remove the two #4-40 retaining screws. 1.008 DISP 1.008 DISP Grasp the bottom cover and lift gently at the 1.008 DISP INPUT JACK END, and remove bottom cover. 1.008 DISP 1.008 DISP Reconnect the battery. 1.009 DISP *NOTE* To read the display with the cover removed, some 1.009 DISP light downward pressure at the top of the exposed PC 1.009 DISP board (A/D) may be required in order to make contact 1.009 DISP through the strip connector between the PCB and LCD. 1.009 DISP 1.009 DISP *NOTE* Handle the PC boards by the edges only. Body 1.009 DISP oil, dirt, etc. can degrade the circuits performance. 1.010 HEAD DC VOLTAGE 1.011 DISP Turn on the UUT. 1.011 DISP *NOTICE* Replace battery if "BAT" is illuminated 1.012 RNG 2V 1.013 PIC 135v 1.014 5500 1.9000V S 2W 1.015 PIC 135adj1 1.016 RNG 20V 1.017 5500 19.000V S 2W 1.018 PIC 135adj2 1.019 5500 * S 1.020 HEAD AC VOLTAGE 1.021 RNG 2V AC 1.022 5500 1.0000V 1kH SI S 2W 1.023 PIC 135adj3 1.024 5500 * S 1.025 HEAD CALIBRATION ADJUSTMENT IS NOW COMPLETE 1.026 DISP Remove all test leads and turn POWER OFF. 1.027 HEAD REASSEMBLY INSTRUCTIONS 1.028 DISP Place the unit face down and remove the battery. 1.028 DISP 1.028 DISP Feed the battery connector through the access opening 1.028 DISP in the bottom of the battery compartment. 1.028 DISP 1.028 DISP Hinge the top and bottom covers together at the 1.028 DISP DIPLAY END. Make sure the TWO catches are aligned in 1.028 DISP in the slots in the bottom cover. 1.029 DISP WARNING 1.029 DISP Be sure to replace the ON-OFF switch cover(if necessary) 1.029 DISP before replacing the bottom cover. A possible hazard 1.029 DISP may exist if cover is NOT replaced. 1.029 DISP 1.029 DISP Install the two #4-40 screws. 1.029 DISP 1.029 DISP Reconnect the battery and replace compartment cover. 1.030 END